masthead_squiggle

Gator® Flex SA Kit – A Reactivable Streptavidin Probe Kit

Introduction

Many of the label-free biosensors have been regenerated successfully to reduce cost for drug research and discovery. However, extensive application of streptavidin biosensors is still hindered by the irreversible immobilization of the bait molecule, and the ability to use the same probe for different application is desirable for cost effective research.

Gator Flex SA Kit

Gator Bio’s Gator Flex SA Kit is the first in the market to provide reactivable streptavidin probes with a reagent set. One probe can easily be used by different users with different applications, which makes it very useful for academics or for use in fully automated screening of different biomolecular interactions. The Gator™ platform enables researcher to monitor whole reactivation process rapidly and provides flexibility and confidence.

Performance Summary

  • Remove both capture reagent and biotin-bait molecule and immobilize new capture reagent < 5 minutes 
  • Use with same or different biotinylated bait 
  • Can be re-used 10 times or more without loss of performance 
  • Automated reactivation 
  • Reagents can be stored at 4°C for up to 5 months 

Results

Loading height reproducibility

Three Flex SA probes were tested over 10 reactivations using biotinylated TNF-α at 1µg/mL. The table below shows CV < 5.0% suggesting excellent loading height reproducibility.

Reproducibility of KD over 10 reactivations using the same bait 

Kinetics was evaluated using biotinylated TNFα and anti-TNFα antibody using a 5 min 1000 rpm pre-wet in K buffer and capture of Flex SA Capture Reagent (SA-CR). The biotinylated TNF-α was loaded onto the sensor, then exposed to association and dissociation of 10, 30, 100 and 300 nM anti TNF-α antibody. Global-fit analysis using Gator Bio software for the TNF-α binding interaction with anti- TNF-α resulted in KD = 7.63E-10 M. 

KD measurements for different proteins over 10 reactivations 

The same probe was tested with 2 different proteins over 10 reactivations. A CV < 5% was observed for both proteins. 

Gator Bio Total Solution 

GatorTM Flex SA kit

 

GatorTM Prime or GatorTM Plus 

Gator Bio Software 

Ordering info 

PL168-350001: GatorTM Flex SA Kit 

Share this article